You are now viewing the owner's manual for your Fisher & Paykel Cooktop, model number .This guide is designed to help you understand the features, functionalities, and maintenance requirements of your Cooktop. Remember, your Fisher & Paykel Cooktop is not just an appliance; it's an investment in your daily comfort and convenience. Please take a moment to familiarize yourself with this manual to ensure you get the most out of your Fisher & Paykel Cooktop.

background image

SPECIFICATIONS 

Dimensions: 

140 mm x 173 mm x 135 mm (without chamber fitted) 

Weight: 

2.8 kg (no chamber fitted), 3.1 kg (chamber fitted and filled with water)

 

Supply Frequency: 

50/60 Hz  Supply Voltage:  115 V~ 

Supply Current:  2.0 A max at 115 V~ 

Heater Plate: 

150 W 

Heater Wire: 

22 V~, 2.73 A, 60 W, 50/60 Hz 

Heater Plate Over-temperature Cutout:  118 ± 6 °C

 

Temperature Control Settings 
Invasive Mode: 

Chamber outlet: 35.5 - 37 °C, Airway 35 - 40 °C 

Noninvasive Mode:  

Chamber outlet: 31 °C, Airway 28 - 34 °C  

Display: 

Three digit 14 mm 7 segment LED 

Range: 

10 - 70 °C, Accuracy: ± 0.3 °C (in 25 - 45 °C temperature range)

 

Alarm Parameters  
High Humidity Alarm: 

An immediate, audible alarm at a displayed temperature of 41 °C or if the  
airway temperature exceeds 43 °C 

Low Humidity Alarm: 

An audible alarm between 10 minutes @ 29.5 °C, and 60 minutes @ 34.5 °C 
(Invasive Mode only) 

 

Sound Pressure Level:  Alarms exceed 50 dBA @ 1 m

 

Performance  
Recommended ambient temperature range:  18 - 26 °C  
CAUTION:  If operating in ambient temperatures outside the recommended range, consult your local 
Fisher & Paykel Healthcare representative. 
Recommended Flow Range:  

Invasive Mode: up to 60 L/min, Noninvasive Mode: up to 120 L/min 

 

Refer to breathing circuit specifications for minimum flow 

Humidity Performance: 

Invasive Mode: > 33 mg/L, Noninvasive Mode: >10 mg/L  

Maximum Operating Pressure:   Refer to chamber and breathing circuit specifications 
Warm-up time:   

Less than 30 minutes 

Symbol Definitions 

   

    

    

    

 

Type BF 

ATTENTION 

Consult 

accompanying 

documents 

Alternating 

Current 

Drip 

Proof 

CAUTION 

Hot surfaces 

may exceed  

85 °C 

Respiratory Humidifier Classified by 

Underwriters Laboratories Inc. with respect to 

electric shock, fire and mechanical hazards 

only in accordance with UL 60601-1 and 

CAN/CSA C22.2 No. 601.1  84DJ 

CLEANING  

MR850 Heaterbase 

 Using a damp cloth, clean the humidifier with either of the following:  Isopropyl Alcohol, normal dishwashing detergent.  Temperature 

Probe 

 Always clean before use.  Probes can be cleaned with one of the following solutions: Sporicidin

TM

, Cidex

TM

 OPA, Sporox

TM

  OR  sterilize the probe using Ethylene Oxide 

sterilization at 55 °C (131 °F), 80 kPa, (allow at least 15 hours for residual ETO to disperse before use).  Wipe the airway temperature probes clear of any cleaning residues 
before use.  Store probes in clean conditions.  DO NOT autoclave probes. 
CAUTION:  DO NOT immerse the heaterbase or temperature probe electrical connections in any liquid. 

 

NOTE: Follow the cleaning agent manufacturers’ instructions carefully. It is the user’s responsibility to qualify any deviations from these procedures, both for disinfecting 
efficacy and physical effect on the probe and the heaterbase. See the MR850 technical manual for more detailed instructions on cleaning. 

WARNING 

Ensure that invasive mode is set for patients 
that have bypassed airways. 

The  use  of  breathing  circuits,  chambers  or 
other  accessories  which  are  not  approved 
by  Fisher  &  Paykel  Healthcare  may  impair 
performance or compromise safety. 

Ensure that both temperature probe sensors 
are  correctly  and  securely  fitted.  Failure  to 
do  so  may  result  in  temperatures  in  excess 
of 41 °C being delivered to the patient. 

Ensure  maintenance  of  grounding  integrity 
by  connection  to  a  "hospital  grade" 
receptacle. Always disconnect supply before 
servicing. 

When  mounting  a  humidifier  adjacent  to  a 
patient  ensure  that  the  humidifier  is  always 
positioned lower than the patient. 

The  operation  of  high  frequency  surgical 
apparatus, 

shortwave 

or 

microwave 

equipment  in  the  vicinity  of  the  humidifier 
may  adversely  affect  its  function.  If  this 
occurs,  the  humidifier  should  be  removed 
from the vicinity of such devices. 

Do  not  touch  the  glass  tip  of  the  chamber 
temperature  probe  during  use.  Keep  black 
connectors dry at all times. 

Visually  inspect  accessories  for 
damage before use. 

Patents Pending in Principal Countries 

of the World 

 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

INSTRUCTION SHEET 

DO NOT DISCARD 

 

REF 185042343  

Rev G 2004-02 

MR850 RESPIRATORY HUMIDIFIER

 

The MR850 respiratory 
humidifier is used to 
warm and humidify 
gases delivered to 
patients requiring 
mechanical ventilation, 
positive pressure 
breathing assistance,  
or other medical gases.

 